CL-NX [GAP Slice Level] Miho 15 May 2026 11:21 Updated Set the Gap sensor slice level. The setting procedure is as follows: 1. Use the following formula to calculate the slice level: (High level - Low level) x 0.3 + Low level = slice level 2. Select the[GAP Slice Level]in the[Calibrate]menu and press thebutton. 3. Press the/buttons to change the[Slice level]value. Set the[Slice level]to the level calculated in step 1. 4. Press thebutton to confirm the value. • If you set the[Slice level]to 0.0 (V), the product sets the slice level automatically. [03-01417-EN]